Limiting Oxygen Index Chamber

The Limiting Oxygen Index (LOI) fire test is perhaps the most economical and precise quality control

WhatsApp: +86 13816217984

Email: info@qinsun-lab.com

Nội dung

The Limiting Oxygen Index (LOI) fire test is perhaps the most economical and precise quality control test for combustible materials. Its ease of use together with high levels of precision has made this technique a primary characterization and quality control tool for the plastic materials and electric cable industries. It is often used in materials specifications and material/product data sheets . The test assesses the minimum percentage of oxygen in the test atmosphere that is required to marginally support combustion.

The unit has been designed to be compact for efficient use in a standard fume cupboard (or under a simple ventilation hood that we can also supply if required). The unit gives continuous digital readout of oxygen concentrations in the test atmosphere to facilitate quick settings of test concentration. Stabilised oxygen percentages are read from the digital readout and no additional flow adjustments are required. This is a considerable improvement over systems that use analogue gauges or require flow matching and the use of graphs or tables to calculate oxygen concentrations. This makes the unit as automatic as is feasible.

While many designs of the LOI test apparatus calculate oxygen index based on the flow of oxygen and nitrogen, the unit incorporates a paramagnetic oxygen analyzer for precise measurement of oxygen concentration without the need for calculations and frequent recalibrations.

image.png
Features

  • Paramagnetic Oxygen Analyzer for assessing accurate oxygen (± 0.1%) levels.
  • Glass test chimney: 450 mm x 75 mm
  • Igniter wand
  • Automatic flow control gives oxygen level adjustment by turning one single valve.
  • Digital display of oxygen percentage in atmosphere during test (no calculations needed).
  • Digital display of temperature of gas mixture entering the test chimney.
  • Quick loading of sample into test chimney.
  • Sample holders for both rigid and flexible samples supplied.
  • One pack of 10 clear PMMA calibration samples (Oxygen Index: 17.4) supplied.


Specification

  • Digital readout for oxygen concentration: +/- 0.1%
  • Dimensions (mm): 350(W) x 370(D) x 280(H)
  • Column size (mm): 75 & 100(dia) x 450(H)
  • Weight (kg): 14
  • Voltage: 230V 50Hz 1A or 110V 60Hz 2A
  • Operating Manual: Supplied
  • Other Service requirements: Bottled O2/N2, Propane

Từ chối trách nhiệm: Nội dung trên chỉ dành cho tham khảo và giao tiếp giữa những người trong ngành và không đảm bảo tính chính xác hoặc đầy đủ của nó. Theo luật pháp và quy định có liên quan và các quy định của trang web này, các đơn vị hoặc cá nhân mua các mặt hàng liên quan phải có trình độ và điều kiện trình độ hợp lệ.