Determines the minimum ignition temperature (MIT) of dust clouds.
WhatsApp: +86 13816217984
Email: info@qinsun-lab.com
Description:
Dust Cloud Minimum Ignition Temperature Tester is mainly used to assess the explosiveness of dust in the production and processing places such as aluminum powder, zinc powder, plastic powder, intermediates of organic synthetic pharmaceuticals, wheat flour, sugar, wood chips, dyestuffs, glued wood ash, milk powder, tea powder, tobacco powder, coal dust, plant fiber dust and so on, and to accurately determine the minimum ignition temperature of the dust cloud.
Reference Standards
ASTM E1491 - 06 EN 50281 - 2 - 1 GB/T 16429 IEC 61241 - 2 ISO/IEC 80079 - 20 - 2
Product Features
1. It has built - in commonly used international test standards, allowing for testing according to standard parameters or custom - setting.
2. It adopts a fuzzy PID temperature - control algorithm to achieve high - precision temperature control.
3. It is equipped with a temperature sensor calibration interface for easy calibration and correction.
4. It uses a Godbert - Greenwald furnace with a built - in quartz tube to ensure the reliable operation of the heating furnace body.
5. The furnace cavity is made of heat - insulating vacuum ceramic fiber, which has a good heat - insulation effect.
6. The temperature thermocouple is from an internationally well - known brand, ensuring the long - term stable operation of the instrument.
7. It has the function of detecting ambient temperature and humidity and automatically records experimental environment information.
8. Remote control ensures the personal safety of experimenters.
9. User - level management enables the management and control of accounts at different levels.
10. The USB interface is plug - and - play, facilitating data management.
11. With professional industrial design, it is simple and elegant.
12. It is equipped with a 7 - inch color liquid - crystal touch screen that dynamically displays the experimental progress.
Technical parameters:
Operating environment | (5 to 45)°C, <85% RH |
Temperature range | Room temperature to 1000°C |
Rising time | ≤20min (room temperature to 500°C) |
Temperature control stability | ±5.0°C |
Temperature measurement accuracy | ±3.0°C |
Temperature measurement precision | ±0.1°C |
Powder spray pressure range | (0 to 100)kPa |
Pressure accuracy | ±5% |
Storage volume | 500mL |
Ambient temperature measurement range | (-10 to 60)°C |
Ambient humidity measurement range | (5 to 95)% RH |
Interface | USB*2 |
Power | 220VAC/50Hz |
Power | 2600W |
Dimensions | 670mm х 400mm х 590mm |
Từ chối trách nhiệm: Nội dung trên chỉ dành cho tham khảo và giao tiếp giữa những người trong ngành và không đảm bảo tính chính xác hoặc đầy đủ của nó. Theo luật pháp và quy định có liên quan và các quy định của trang web này, các đơn vị hoặc cá nhân mua các mặt hàng liên quan phải có trình độ và điều kiện trình độ hợp lệ.
Điện thoại công ty
+86-21-6420 0566
Thời gian làm việc
Thứ Hai đến Thứ Sáu
Điện thoại di động:
13816217984
Email:
info@qinsun-lab.com