Product introduction The anchoring force test of the guide needle can ensure the anchoring effect o
WhatsApp: +86 13816217984
Email: info@qinsun-lab.com
Product introduction
The anchoring force test of the guide needle can ensure the anchoring effect of the mammary positioning filament within the tissue and prevent the filament from shifting or falling out before surgery. This is essential for stable positioning filaments to accurately locate and remove the tumor during surgery.
Test standard
YY/T0880-2013 "Disposable Breast Positioning wire and Guide Needle"
Technical parameter
1. Tensile force test range: 0——20N, error is not more than +1%, resolution 0.01N
2. Test time: 1-999S, any choice, the error is not more than +0.5S
3. Test speed: 10mm/min, 50mm/min, 100m/min
4. Distance adjustment: 10mm——50mm adjustable
5. Sensor: high-precision force value sensor (with hardware and software protection function)
6. Fixture: high quality 304 stainless steel
7. Force measuring mechanism: built-in
8. Printer: on-board micro needle printer
9. Data storage: USB disk can be imported into the computer storage
10. Machine power: 70W
11. Power supply: AC90V-240V/50Hz
Từ chối trách nhiệm: Nội dung trên chỉ dành cho tham khảo và giao tiếp giữa những người trong ngành và không đảm bảo tính chính xác hoặc đầy đủ của nó. Theo luật pháp và quy định có liên quan và các quy định của trang web này, các đơn vị hoặc cá nhân mua các mặt hàng liên quan phải có trình độ và điều kiện trình độ hợp lệ.